Wenande E, Hastrup A, Fredman GL, Olesen UH, Wolkerstorfer A, Haedersdal M. Lasers Surg Med. 2024 Jan;56(1):81-89.

Impact of skin hydration on patterns of microthermal injury produced by fractional CO2 laser

Abstract

Objectives: The impact of skin hydration on patterns of thermal injury produced by ablative fractional lasers (AFLs) is insufficiently examined under standardized conditions. Using skin with three different hydration levels, this study assessed the effect of hydration status on microchannel dimensions generated by a fractional CO2 laser.

Methods: A hydration model (hyperhydrated-, dehydrated- and control) was established in ex vivo porcine skin, validated by changes in surface conductance and sample mass. After, samples underwent AFL exposure using a CO2 laser (10,600 nm) at two examined pulse energies (10 and 30 mJ/mb, fixed 10% density, six repetitions per group). Histological assessment of distinct microchannels (n = 60) determined three standardized endpoints in H&E sections: (1) depth of microthermal treatment zones (MTZs), (2) depth of microscopic ablation zones (MAZs), and (3) coagulation zone (CZ) thickness. As a supplemental in vivo assessment, the same laser settings were applied to hyperhydrated- (7-h occlusion) and normohydrated forearm skin (no pretreatment) of a human volunteer. Blinded measurement of MAZ depth (n = 30) was performed using noninvasive optical coherence tomography (OCT).

Results: Modest differences in microchannel dimensions were shown between hyperhydrated, dehydrated and control skin at both high and low pulse energy. Compared to controls, hyperhydration led to median reductions in MTZ and MAZ depth ranging from 5% to 8% (control vs. hyperhydrated at 30 mJ/mb; 848 vs. 797 µm (p < 0.003) (MAZ); 928 vs. 856 µm (p < 0.003) (MTZ)), while 14%-16% reductions were shown in dehydrated skin (control vs. dehydrated at 30 mJ/mb; MAZ: 848 vs. 727 µm (p < 0.003); MTZ: 928 vs. 782 µm (p < 0.003)). The impact of skin hydration on CZ thickness was in contrast limited. Corresponding with ex vivo findings, hyperhydration was similarly associated with lower ablative depth in vivo skin. Thus, median MAZ depth in hydrated skin was 10% and 14% lower than in control areas at 10 and 30 mJ/mb pulse energy, respectively (10 mJ: 210 vs. 180 µm (p < 0.001); 30 mJ: 335 vs. 300 µm (p < 0.001)).

Conclusion: Skin hydration status can exert a minimal impact on patterns of microthermal injury produced by fractional CO2 lasers, although the clinical implication in the context of laser therapy requires further study.